New ATONA Instrument Data
As the new Phoenix instruments with the ATONA amplifiers move through production we have a lot of new data to share. Here we are running 5ng loads of NIST SRM 987 from a recent Phoenix installation. The average internal precision is 3.6ppm while the external is 5.68ppm.
87/86 2 rse (ppm)
0.710256 3.7
0.71026 3.6
0.710257 2.5
0.710253 3.5
0.710252 4.7
0.710252 3.8
0.710251 3.7
0.710247 3.5
Mean 0.7102535
Std Dev 4.03556E-06
1RSD (ppm) 5.68ppm
We also have some very exciting data showing the analysis of Calcium. For a 1000ng load, the average internal precision for 40Ca/44Ca is 14ppm 1RSE,while the external precision is 25.1ppm 1RSD.
40/44 RSE (ppm)
47.1109 14
47.1109 14
47.109 17
47.1082 14
47.1098 12
Mean 47.1097
Std Dev 0.00118
RSD ppm 25.1
Please stop by and see us at the Goldschmidt 2018 meeting in Boston (booth 43) we will have lots more data to share.